Humidity calibration set EE-HA010401, EE-HA010402, EE-HA010435, EE-HA010480

70The most simple solution for the periodic check and calibration of humidity transmitters is the E+E calibration set. This set consists of humidity standards and different calibration devices, carefully designed for use with E+E humidity sensor probes.
Humidity calibration set 
The E+E humidity calibration set is a convenient way to expose sensor to RH of 5, 10, 35, 50, 80 & 95% for checking or calibration. Solutions come in sets of 5 glass ampoules.
  • Versions for horizontal or vertical use
  • Convenient & simple
  • Accuracy ±0.5% at 23 ±2°C
  • Non-saturated lithium chloride solution

The air above the salt solutions will hold a fixed RH at a certain temperature which is not STRONGLY temperature dependent. High accuracy calibration can be attained if the temperature is held very constant and the variation from the reference temperature is allowed for. The sensor is isolated from the liquid by a semi permeable membrane.

E+E calibration set guarantees easy and reliable instrument check


Get More InformationGet Quote

Parts Listing

EE-HA010401CALIBRATION DEVICE Horizontal Not In Stock
EE-HA010435RH STANDARD 35% 5Ampoules/Pads Not In Stock
EE-HA010480RH STANDARD 80% 5Ampoules/Pads Not In Stock
EE-HA010402CALIBRATION DEVICE Vertical Not In Stock

Document Listing

SpecificationHumidity Calibration Set